Mitigation of Blood Borne Cell Attachment to Metal Implants through CD47-Derived Peptide Immobilization
Published on: December 3, 2020
Mitigation effect of cell exclusion on blood damage in spiral groove bearings
Chris Hoi Houng Chan1, Tomotaka Murashige2, Shelby A Bieritz3
1School of Engineering and Built Environment, Griffith University, Queensland, Australia; Critical Care Research Group, The Prince Charles Hospital, Brisbane, Australia.
Cell exclusion in spiral groove bearing (SGB) excludes red blood cells from high shear regions in the bearing gaps and potentially reduce haemolysis in rotary blood pumps. However, this mechanobiological phenomenon has been observed in ultra-low blood haematocrit only, whether it can mitigate blood damage in a clinically-relevant blood haematocrit remains unknown. This study examined whether cell exclusion in a SGB alters haemolysis and/or high-molecular-weight von Willebrand factor (HMW vWF) multimer degradation. Citrated human blood was adjusted to 35 % haematocrit and exposed to a SGB (n = 6) and grooveless disc (n = 3, as a non-cell exclusion control) incorporated into a custom-built Couette test rig operating at 2000RPM for an hour; shearing gaps were 20, 30, and 40 μm. Haemolysis was assessed via spectrophotometry and HMW vWF multimer degradation was detected with gel electrophoresis and immunoblotting. Haemolysis caused by the SGB at gaps of 20, 30 and 40 μm were 10.6 ± 3.3, 9.6 ± 2.7 and 10.5 ± 3.9 mg/dL.hr compared to 23.3 ± 2.6, 12.8 ± 3.2, 9.8 ± 1.8 mg/dL.hr by grooveless disc. At the same shearing gap of 20 µm, there was a significant reduced in haemolysis (P = 0.0001) and better preserved in HMW vWF multimers (p < 0.05) when compared SGB to grooveless disc. The reduction in blood damage in the SGB compared to grooveless disc is indicative of cell exclusion occurred at the gap of 20 µm. This is the first experimental study to demonstrate that cell exclusion in a SGB mitigates the shear-induced blood damage in a clinically-relevant blood haematocrit of 35 %, which can be potentially utilised in future blood pump design.
Cell exclusion in spiral groove bearing (SGB) excludes red blood cells from high shear regions in the bearing gaps and potentially reduce haemolysis in rotary blood pumps. However, this mechanobiological phenomenon has been observed in ultra-low blood haematocrit only, whether it can mitigate blood damage in a clinically-relevant blood haematocrit remains unknown. This study examined whether cell exclusion in a SGB alters haemolysis and/or high-molecular-weight von Willebrand factor (HMW vWF) multimer degradation. Citrated human blood was adjusted to 35 % haematocrit and exposed to a SGB (n = 6) and grooveless disc (n = 3, as a non-cell exclusion control) incorporated into a custom-built Couette test rig operating at 2000RPM for an hour; shearing gaps were 20, 30, and 40 μm. Haemolysis was assessed via spectrophotometry and HMW vWF multimer degradation was detected with gel electrophoresis and immunoblotting. Haemolysis caused by the SGB at gaps of 20, 30 and 40 μm were 10.6 ± 3.3, 9.6 ± 2.7 and 10.5 ± 3.9 mg/dL.hr compared to 23.3 ± 2.6, 12.8 ± 3.2, 9.8 ± 1.8 mg/dL.hr by grooveless disc. At the same shearing gap of 20 µm, there was a significant reduced in haemolysis (P = 0.0001) and better preserved in HMW vWF multimers (p < 0.05) when compared SGB to grooveless disc. The reduction in blood damage in the SGB compared to grooveless disc is indicative of cell exclusion occurred at the gap of 20 µm. This is the first experimental study to demonstrate that cell exclusion in a SGB mitigates the shear-induced blood damage in a clinically-relevant blood haematocrit of 35 %, which can be potentially utilised in future blood pump design.
Related Concept Videos
Collar Bearings
Bearings: Problem Solving
Bearing Stress
Due to the intricacy of these microforces, an average value, known as bearing stress, is often used by...
Stress Concentrations in Circular Shafts
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Pivot Bearings
A pivot bearing is a specialized type of bearing designed to support axial loads on a rotating shaft. The bearing surface, or the pivot, is positioned at the end of a shaft to support the axial thrust. The pivot may...