Improving the accuracy of copolymer sequence length determination by pyrolysis gas chromatography: A comprehensive
Wouter C Knol1, Sifra Vos1, Till Gruendling2
1Analytical Chemistry Group, van 't Hoff Institute for Molecular Sciences (HIMS), Faculty of Science, University of Amsterdam, Science Park 904, Amsterdam, the Netherlands; Centre for Analytical Sciences Amsterdam, Science Park 904, Amsterdam, the Netherlands.
Abstract:
Many industrial polymers which find application in contemporary materials are copolymers. Copolymers feature multiple distributions, that govern their physical properties, including the sequence distribution. Styrene-acrylate copolymers are an important class of polymers, their monomer sequence is typically determined by 13C NMR which suffers from low sensitivity and spectral resolution. A series of studies have shown that Py-GC can be applied to determine the sequence length of copolymers. The accuracy of the trimer assignments and the appropriate calibration approaches yielding reliable data have however not yet been validated. In the present study we propose a comprehensive workflow to ensure the accuracy of the sequence determination by Py-GC, next to NMR. In-depth analysis of the trimers observed in the Py-GC pyrograms of model styrene-acrylate copolymers was performed and specific MS fragments relating to the trimer sequence were assigned. A comparison of a series of copolymers yielded reliable assignments for the trimer signals. The obtained sequence lengths were in agreement with those calibrated with the benchmark method, 13C NMR. Py-GC was found to consistently underestimate the acrylate sequence length. Py-GC calibration with 13C NMR was thus found to be indispensable for the accurate absolute quantification of the sequence length by Py-GC. The calculated randomness did not vary significantly after NMR calibration, indicating that NMR calibration might not be required in all cases to obtain (relative) information on the sequence of a copolymer.
More Related Videos
09:34Improved Polymerase Chain Reaction-restriction Fragment Length Polymorphism Genotyping of Toxic Pufferfish by Liquid Chromatography/Mass Spectrometry
Published on: September 20, 2016
06:56Characterization of Synthetic Polymers via Matrix Assisted Laser Desorption Ionization Time of Flight MALDI-TOF Mass Spectrometry
Published on: June 10, 2018
Related Concept Videos
Maxam-Gilbert Sequencing
Challenges of the Maxam-Gilbert Method
The...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...