Related Experiment Video
Updated: Aug 18, 2025

Veno-Venous Extracorporeal Membrane Oxygenation in a Mouse
Published on: October 24, 2018
[Emergency treatment of centrifugal pump dysfunction in extracorporeal membrane oxygenation]
Ke Fang1, Jun Wang, Ruixiang Sun
1Department of Critical Care Medicine, the First Affiliated Hospital of Wannan Medical College (Yijishan Hospital), Anhui Provincial Clinical Research Center for Critical Respiratory Disease, Wuhu 241000, Anhui, China. Corresponding author: Jiang Xiaogan,
Objective:
To summarize the treatment process of rapid decrease in blood flow due to centrifugal pump dysfunction during extracorporeal membrane oxygenation (ECMO) and its related thinking.
Methods:
On September 25, 2021, the ECMO treatment of a 14-year-old boy with severe mycoplasma pneumonia, severe viral pneumonia and acute respiratory distress syndrome (ARDS) admitted to the department of critical care medicine of the First Affiliated Hospital of Wannan Medical College was analyzed.
Results:
Oxygenation of the child was difficult to maintain under invasive mechanical ventilation, and lung consolidation progressed seriously. After evaluation, venous-venous ECMO (VV-ECMO) was implemented, then oxygenation was improved. In the 120th hour after VV-ECMO establishment, the blood flow sudden decreased, the speed was 3 822 r/min, while the flow was only 0.2 L/min, more over there was no change in the flow when the speed was increased. Before that, the ECMO speed was 3 530 r/min, and the flow was up to 3.4 L/min and stable. After rapid screening, it was determined that the centrifugal pump was dysfunction. ECMO was successfully replaced and the flow was satisfactory.
Conclusions:
At present, most ECMO centers do not routinely monitor the pressure before and after the pump. There is a lack of visual and quantitative techniques or indicators to judge the pump's function, and there is also a lack of corresponding clinical experience in treatment. This paper summarizes the investigation and treatment process of ECMO pump dysfunction of this case to provide reference.
Related Concept Videos
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Continuous Renal Replacement Therapy
Heart Failure VI: Adjunct Therapies
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Hemodialysis II: Procedure and Complications