Background enhancement in contrast-enhanced spectral mammography (CESM): are there qualitative and quantitative
Daniel Wessling1, Simon Männlin2, Ricarda Schwarz2
1Department of Diagnostic and Interventional Radiology, University Hospital of Tuebingen, Hoppe-Seyler-Strasse 3, 72076, Tuebingen, Germany. Daniel.wessling@med.uni-tuebingen.de.
Objective:
To evaluate the impact of the digital mammography imaging system on overall background enhancement on recombined contrast-enhanced spectral mammography (CESM) images, the overall background enhancement of two different mammography systems was compared.
Methods:
In a retrospective single-center study, CESM images of n = 129 female patients who underwent CESM between 2016 and 2019 were analyzed independently by two radiologists. Two mammography machines of different manufacturers were compared qualitatively using a Likert-scale from 1 (minimal) to 4 (marked overall background enhancement) and quantitatively by placing a region of interest and measuring the intensity enhancement. Lesion conspicuity was analyzed using a Likert-scale from 1 (lesion not reliably distinguishable) to 5 (excellent lesion conspicuity). A multivariate regression was performed to test for potential biases on the quantitative results.
Results:
Significant differences in qualitative background enhancement measurements between machines A and B were observed for both readers (p = 0.003 and p < 0.001). The quantitative evaluation showed significant differences in background enhancement with an average difference of 75.69 (99%-CI [74.37, 77.02]; p < 0.001). Lesion conspicuity was better for machine A for the first and second reader respectively (p = 0.009 and p < 0.001). The factor machine was the only influencing factor (p < 0.001). The factors contrast agent, breast density, age, and menstrual cycle could be excluded as potential biases.
Conclusion:
Mammography machines seem to significantly influence overall background enhancement qualitatively and quantitatively; thus, an impact on diagnostic accuracy appears possible.
Key Points:
• Overall background enhancement on CESM differs between different vendors qualitatively and quantitatively. • Our retrospective single-center study showed consistent results of the qualitative and quantitative data analysis of overall background enhancement. • Lesion conspicuity is higher in cases of lower background enhancement on CESM.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Magnetic Resonance Imaging
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...