Objective:
Many hybrid closed-loop (HCL) systems struggle to manage unusually high glucose levels as experienced with intercurrent illness or pre-menstrually. Manual correction boluses may be needed, increasing hypoglycemia risk with overcorrection. The Cambridge HCL system includes a user-initiated algorithm intensification mode ("Boost"), activation of which increases automated insulin delivery by approximately 35%, while remaining glucose-responsive. In this analysis, we assessed the safety of "Boost" mode.
Methods:
We retrospectively analyzed data from closed-loop studies involving young children (1-7 years, n = 24), children and adolescents (10-17 years, n = 19), adults (≥24 years, n = 13), and older adults (≥60 years, n = 20) with type 1 diabetes. Outcomes were calculated per participant for days with ≥30 minutes of "Boost" use versus days with no "Boost" use. Participants with <10 "Boost" days were excluded. The main outcome was time spent in hypoglycemia <70 and <54 mg/dL.
Results:
Eight weeks of data for 76 participants were analyzed. There was no difference in time spent <70 and <54 mg/dL between "Boost" days and "non-Boost" days; mean difference: -0.10% (95% confidence interval [CI] -0.28 to 0.07; P = .249) time <70 mg/dL, and 0.03 (-0.04 to 0.09; P = .416) time < 54 mg/dL. Time in significant hyperglycemia >300 mg/dL was 1.39 percentage points (1.01 to 1.77; P < .001) higher on "Boost" days, with higher mean glucose and lower time in target range (P < .001).
Conclusions:
Use of an algorithm intensification mode in HCL therapy is safe across all age groups with type 1 diabetes. The higher time in hyperglycemia observed on "Boost" days suggests that users are more likely to use algorithm intensification on days with extreme hyperglycemic excursions.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Hypoglycemia and Glucagon
Insulin: Biosynthesis, Chemistry, and Preparation
Damage or functional impairment of β-cells inhibits insulin production, leading to diabetes. Diabetes treatment...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...