Bench-Side Dose Accuracy of an Open-Source Ultra-Low-Cost Insulin-Pump, With Testing Conducted to IEC 60601-2-24
Matthew Payne1, Francis Pooke1, Lui Holder-Pearson1,2
1Centre for Bio-engineering, Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand.
Abstract:
With the prevalence of diabetes higher than ever, governments and people with diabetes are facing significant treatment and indirect costs associated with managing their condition. An ultra-low-cost insulin pump is a possible solution to improving health disparities. This article presents test results for an insulin-pump built from low-cost components (bill of materials < $US100). All testing was completed in accordance with IEC60601-2-24, and results were benchmarked against a commercial pump. Results showed the ultra-low-cost pump has comparable accuracy to the commercially available insulin pump with testing displaying an overall accuracy of 0.089% and -0.392%, respectively. These results show that an ultra-low-cost pump can accurately deliver insulin in limited bench testing. Testing in other environments and scenarios is required to fully meet IEC60601-2-24 standards.