The method development for elimination of matrix interferences in seawater monitoring to determine elements by ICP-MS
Murat Kilic1, Serpil Kilic2, Serpil Yenisoy-Karakaş3
1Central Research Laboratory Application and Research Center, Isparta University of Applied Sciences, Isparta, Turkey.
Abstract:
A simple, versatile, and economical method development with matrix elimination to determine the elements As, Ba, Cd, Co, Cr, Cu, Mn, Ni, Pb, V, and Zn in seawater by using the technique of inductively coupled plasma mass spectrometry is reported. Real seawater was used as a matrix for the standard adding calibration and other validation parameters. The samples were open digested at 80 °C with 2 mL HNO3 and 0.25 mL HF under the hood. A two-step digestion procedure was applied, and the volume was completed to 10 mL with deionized water. Chloride removal was accomplished by using this procedure. The concentrations of Ca and Mg ions were lessened by 15% and 20%, respectively. These results were verified with ion chromatography, SEM-EDS, and mass difference analyses. It was observed that there was a 40% loss in the average mass of particulate matter on the filter media after applying the two-step digestion procedure. Recovery and trueness values were in the range of 86 and 109%. The average precision amounts for elements were determined as RSD (%) in the range of 1.0% and 3.4%. The concentrations of elements determined in the 18 samples collected from the Konyaaltı Beach located in Antalya were higher than the maximum allowable concentrations of the Directive 2013/39/EU.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...