Related Experiment Video
Updated: Aug 18, 2025

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Technical note: Water-equivalent thickness of Superflab bolus material at diagnostic x-ray energies
Kayla F Blunt1, Emily L Marshall1, Chao Guo1
1Department of Radiology, The University of Chicago, Chicago, Illinois, USA.
Purpose:
The purpose of this work was to determine the water-equivalent thickness of Superflab bolus material for narrow and broad field-of-view (FOV) x-ray geometries at diagnostic x-ray energies.
Methods:
Transmission measurements were performed for incremental thicknesses of Superflab bolus material and water in narrow and broad FOV x-ray geometries. The transmission data was fit to a non-linear model for x-ray transmission - the Archer model. Water-equivalent thickness of Superflab was calculated based upon fitting parameters to transmission curves for 75, 95, and 115 kV x-ray tube voltages. Measured x-ray transmission factors for water and Superflab were used to determine the water equivalence of Superflab.
Results:
For all x-ray tube voltages and geometries, the water equivalence of Superflab was greater than one, indicating that Superflab is more attenuating than water. This effect was stronger for broad FOV geometries. At 95 kV, 30 cm of Superflab corresponded to 32.0 cm of water in the narrow FOV geometry, and 34.3 cm of water in the broad FOV geometry. The Archer model fitting parameters and Superflab water equivalence are reported for all x-ray beam conditions explored in this work.
Conclusions:
Superflab bolus material is more attenuating than water at diagnostic x-ray energies. The Archer model and its respective fitting parameters reported in this work may be used to estimate the water-equivalent thickness of Superflab for diagnostic x-ray spectra.
Related Concept Videos
X-ray Imaging
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Upper GI Series: Barium Swallow
Purpose and Procedure
Patients undergoing this procedure ingest a liquid containing barium sulfate with a chalky...
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...