Related Experiment Video
Updated: Aug 18, 2025

MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
A Comparison of Clinical and Radiological Outcomes Between Target 360 Nano and Microplex Hypersoft 3D Used as
Sang Uk Kim1, Joon Huh, Dal Soo Kim
1Myong-Ji St. Mary?s Hospital, Department of Neurosurgery, Seoul, South Korea.
Aim:
To compare the clinical outcomes of Target 360 nano (TG) and Microplex hypersoft 3D (MH) used as a finishing coil (FC).
Material And Methods:
From January 2018 to December 2020, we retrospectively reviewed 243 coil embolization procedures performed using TG (n=152) and MH (n=91) coils of 1mm x 2 cm the same size as FC. Further, the clinical and radiographic results were compared by matching the propensity score between the two groups.
Results:
There were no statistically significant differences in the clinical and angiographic results of the two coils after the propensity score matching. Successful occlusion was 89% and 86.8% and FC insertion failure was 20.9% and 28.6%. There were no differences in procedure-related complications and recurrence between the groups during the eight months follow-up period (3.3% versus 4.4% and 4.4% versus 3.3%, respectively). We also compared two subgroups of failed FC insertion (19 of TG and 26 of MH). The number of angled catheters was significantly higher in the failed TG group than in the failed MH group.
Conclusion:
There was no statistically significant difference between the clinical and radiological outcomes of TG and MH used as FC. However, in the FC insertion failure subgroups, the number of angled catheters was significantly higher in the TG failed group than in the MH failed. It was experimentally confirmed that the angle change of microcatheter tip with a large angle was large; however, further studies are required.
More Related Videos
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
07:29Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...