Related Experiment Video
Updated: Aug 18, 2025

Coincubation Assay for Quantifying Competitive Interactions between Vibrio fischeri Isolates
Published on: July 22, 2019
How to assess parallelism in factor assays: coefficient of variation of results with different dilutions or slope
Susan Guy1, M Fiona Shepherd1, Annette E Bowyer1
1Sheffield Haemophilia and Thrombosis Centre, Royal Hallamshire Hospital, Sheffield, UK.
Introduction:
Non-parallelism in factor assays can lead to incorrect factor activities. Parallelism can be assessed by calculating the coefficient of variation (CV) of results obtained on 3 dilutions of the same sample. Some authors have proposed that if there is <15% then the average activity is reportable. Some analysers use a slope ratio (SR) to calculate parallelism, with an acceptance range of approximately 0.9-1.1.
Methods:
We evaluated CV and SR in one stage FII-FXII assays on Sysmex CS5100i using Innovin or Actin FS. Frozen normal and pathological plasmas, plasmas containing Direct Oral Anticoagulants, Direct Thrombin Inhibitors or Lupus Anticoagulant were analysed to assess possible non-parallelism.
Results:
In plasmas with factor levels >25 IU/dl (plus no interfering substances) all CVs were < 15%. One sample (low factor activities 10-15 IU/dl), had CVs > 15% in FII, FVII and FXII assays only. SR outside of 0.9-1.1 were seen in FII and FXII assays at different levels of clotting factor including some within the normal range. Non-parallelism was detected more frequently with SR than CV for those with interfering substances.
Conclusions:
SR outside of 0.9-1.1 were seen in different levels of clotting factors, including samples which did not contain interfering substances. The target of 15% CV was a better discriminator than a SR for acceptance. When factor levels were reduced to around 10-15 IU/dl, a target 20 %CV was more appropriate than 15%. It might be appropriate for laboratories to assess locally whether their acceptance criteria need to be wider at low levels of clotting factors.
Related Concept Videos
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
Coefficient of Variation
The coefficient of variation is a practical statistical tool in finance. It allows investors to assess the volatility or...
Factors Affecting Activity Coefficient
The activity coefficient value for an ion is close to one when the solution has almost zero ionic strength, i.e., when the solution shows close to ideal behavior. As the ionic strength of the solution increases from 0 to 0.1 mol/L, a...