A Comparative Study of Forehead Temperature and Core Body Temperature under Varying Ambient Temperature Conditions
Anming Chen1,2, Jia Zhu1,2,3, Qunxiong Lin4
1Tsinghua Shenzhen International Graduate School, Tsinghua University, Shenzhen 518055, China.
Abstract:
When the ambient temperature, in which a person is situated, fluctuates, the body's surface temperature will alter proportionally. However, the body's core temperature will remain relatively steady. Consequently, using body surface temperature to characterize the core body temperature of the human body in varied situations is still highly inaccurate. This research aims to investigate and establish the link between human body surface temperature and core body temperature in a variety of ambient conditions, as well as the associated conversion curves.
Methods:
Plan an experiment to measure temperature over a thousand times in order to get the corresponding data for human forehead, axillary, and oral temperatures at varying ambient temperatures (14-32 °C). Utilize the axillary and oral temperatures as the core body temperature standards or the control group to investigate the new approach's accuracy, sensitivity, and specificity for detecting fever/non-fever conditions and the forehead temperature as the experimental group. Analyze the statistical connection, data correlation, and agreement between the forehead temperature and the core body temperature.
Results:
A total of 1080 tests measuring body temperature were conducted on healthy adults. The average axillary temperature was (36.7 ± 0.41) °C, the average oral temperature was (36.7 ± 0.33) °C, and the average forehead temperature was (36.2 ± 0.30) °C as a result of the shift in ambient temperature. The forehead temperature was 0.5 °C lower than the average of the axillary and oral temperatures. The Pearson correlation coefficient between axillary and oral temperatures was 0.41 (95% CI, 0.28-0.52), between axillary and forehead temperatures was 0.07 (95% CI, -0.07-0.22), and between oral and forehead temperatures was 0.26 (95% CI, 0.11-0.39). The mean differences between the axillary temperature and the oral temperature, the oral temperature and the forehead temperature, and the axillary temperature and the forehead temperature were -0.08 °C, 0.49 °C, and 0.42 °C, respectively, according to a Bland-Altman analysis. Finally, the regression analysis revealed that there was a linear association between the axillary temperature and the forehead temperature, as well as the oral temperature and the forehead temperature due to the change in ambient temperature.
Conclusion:
The changes in ambient temperature have a substantial impact on the temperature of the forehead. There are significant differences between the forehead and axillary temperatures, as well as the forehead and oral temperatures, when the ambient temperature is low. As the ambient temperature rises, the forehead temperature tends to progressively converge with the axillary and oral temperatures. In clinical or daily applications, it is not advised to utilize the forehead temperature derived from an uncorrected infrared thermometer as the foundation for a body temperature screening in public venues such as hospital outpatient clinics, shopping malls, airports, and train stations.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Factors Affecting Body Temperature
Factors may include: