Real-time phase contrast MRI versus conventional phase contrast MRI at different spatial resolutions and velocity
Pan Liu1, Sidy Fall2, Maureen Ahiatsi3
1CHIMERE UR 7516, Jules Verne University of Picardy, Amiens, France; Medical Image Processing Department, Amiens Picardy University Hospital, Amiens, France.
Purposes:
To compare the accuracy of real-time phase-contrast echo-planar MRI (EPI-PC) and conventional cine phase-contrast MRI (Conv-PC) and to assess the influence of spatial resolutions (pixel size) and velocity encoding on flow measurements obtained with the two sequences.
Methods:
Flow quantification was assessed using a pulsatile flow phantom (diameter: 9.5 mm; mean flow rate: 1150 mm3/s; mean flow velocity: 1.6 cm/s). Firstly, the accuracy of the EPI-PC was checked by comparing it with the flow rate in the calibrated phantom and the pulsation index from Conv-PC. Secondly, flow data from the two sequences were compared quantitatively as a function of the pixel size and the velocity encoding.
Results:
The mean percentage difference between the EPI-PC flow rate and calibrated phantom flow rate was -2.9 ± 2.1% (Mean ± SD). The pulsatility indices for EPI-PC and Conv-PC were respectively 0.64 and 0.59. In order to keep the flow rate measurement error within 10%, the ROI in Conv-PC had to contain at least 13 pixels, while the ROI in EPI-PC had to contain at least 9 pixels. Furthermore, Conv-PC had a higher velocity-to-noise ratio and could use a higher velocity encoding than EPI-PC (20 cm/s and 15 cm/s, respectively).
Conclusions:
The result of this in vitro study confirmed the accuracy of EPI-PC, and found that EPI-PC can adapt to lower spatial resolutions, but is more sensitive to velocity encoding than Conv-PC.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...