Continuous glucose monitoring for diabetes: potential pitfalls for the general physician
Samuel S Herrod1, Grace Liversedge1, Bijay Vaidya1
1MacLeod Diabetes and Endocrine Centre, Exeter, UK.
Abstract:
A 31-year-old man presented systemically unwell with diabetic ketoacidosis (DKA). He was using an intermittently scanned continuous glucose monitoring (CGM) device that had been recording low or normal glucose readings for the 48 hours prior to admission. The sensor site had become infected, and we believe this soft tissue infection caused his CGM device to record falsely low glucose readings leading the patient to erroneously lower his insulin doses and take extra carbohydrates, precipitating DKA. CGM devices measure glucose levels in the interstitial fluid. When interstitial glucose readings do not match symptoms or expectations, a capillary blood glucose reading should be taken to correlate and impact treatment decisions. There will be an increase in patients presenting to hospital with CGM devices as the National Institute for Health and Care Excellence guidelines have recently been updated. We use this interesting clinical case to provide context for key learning points about CGM devices for the general physician.
More Related Videos
Related Concept Videos
Hypoglycemia and Glucagon
Diabetes: Symptoms, Diagnosis, and Complications
Diabetes Mellitus: Type 2 and Gestational
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Errors occurring during blood pressure monitoring
Several factors...