Mechanical comparison of standard interlocking, clawed, and expandable wedge locked nail fixations: An experimental
Mustafa Özkaya1, Selçuk Tunalı2, İsmet Köksal3
1Department of Mechanical Engineering, KTO Karatay University, 42020 Konya, TURKEY.
Introduction:
Intramedullary nailing has been used as a standard in the treatment of the long bone fractures with its clinical and mechanical advantages. However, using distal locking screws has been associated with longer operative times, higher radiation exposure rates, and complications like breakages of distal screw or nail at the screw hole level. Therefore, attempts to eliminate distal locking screws has been always present for intramedullary nail fixation. With a similar purpose, the present study has been carried out to compare mechanical behaviors of intramedullary nail fixations with different distal locking elements.
Materials And Methods:
In this study, mechanical behaviors of standard interlocking and clawed nail fixations were compared experimentally in the first part. Six fourth generation Sawbones femurs, which have a simulated subtrochanteric fracture, were divided equally and were fixed with standard interlocking and clawed nails. During axial compression tests, experimental strain measurements were taken from all fixations. After validation of numerical models with using experimental strains and stiffnesses, mechanical behaviors of standard interlocking, clawed, and wedge locked nail fixations were compared numerically under axial compression loads. In numerical comparison, the stress-strain distributions were evaluated.
Results:
Experimental results showed that although that there was no significant difference in stiffnesses, standard nail fixation bore two times higher loads than clawed nail fixations. Under loading, decrease in the distance between fracture surfaces was approximately seven times higher in the clawed nail fixation when compared to standard nail fixations. Numerical results showed that wedge locked nail fixation provided equivalent mechanical behavior to standard interlocking nail.
Discussion:
In experiments of clawed nail fixation, the high decrease in the distance between fracture surfaces was evidence of the slippage of nail in the medullary canal. For a safe fixation, claws should be deployed when they are completely in contact with the cortical bone, they should be stuck into the bone in a fair amount, and the deployment in the distal third of the femur should be avoided. According to experimentally validated numerical analyses, wedge locked nail fixation may be an alternative for standard interlocking nail fixation if experimental studies support the present results.
Related Concept Videos
Wood Fasteners
Nails are the most common fasteners, consisting of sharp-pointed metal pins that are driven into wood using a hammer or a mechanical nail gun. They come in a variety of diameters, heads, and lengths. Nailing techniques include...
Wedges
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...
Moment of a Force: Problem Solving
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Mortar Properties