Related Experiment Video
Updated: Aug 17, 2025

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Electrocardiographic characteristics for the prediction of under-sensing in implantable loop recorders
Hirota Kida1, Masato Kawasaki2, Yoshitaka Kikuchi1
1Department of Clinical Engineering Osaka General Medical Center Osaka Japan.
Background:
Under-sensing (US) in implantable loop recorders (ILRs) interferes with the accurate diagnosis of arrhythmia, but there is little information available on the details of US of ILRs. The aim of this study was to clarify the frequency of US in patients with ILRs and to investigate the predictors of US in ILRs prior to implantation.
Methods And Results:
We studied 46 consecutive patients implanted ILR. During the mean follow-up period of 499 ± 363 days, 15 events of US were observed in five patients. There were no significant differences in patient characteristics between patients with and without US. In standard 12-lead electrocardiogram (ECG), QRS complex amplitude in anterolateral chest leads (V2 to V5) were significantly lower in patients with than without US (V2: 0.88 [0.66, 1.22] mV vs. 1.67 [1.23, 2.29] mV, p = .010 V3: 1.25 [1.00, 1.26] mV vs. 1.90 [1.41, 2.29] mV, p = .013; V4: 1.14 [0.96, 1.38] mV vs. 1.93 [1.65, 2.64] mV, p = .023; V5: 0.57 [0.50, 0.75] mV vs. 1.60 [1.20, 1.98] mV, p = .011, respectively). ROC curve analysis showed that cut-off values of 1.30 mV of QRS complex amplitude in V2, 1.26 mV of that in V3, and 0.75 mV of that in V5 had moderate accuracy for predicting US (V2: sensitivity 68%, specificity 100%, area under the curve [AUC] 0.86; V3: sensitivity 85%, specificity 80%, AUC 0.85; V5: sensitivity 98%, specificity 80%, AUC 0.85, respectively).
Conclusions:
US was observed in 10.9% patients with an ILR. QRS complex amplitude in anterolateral chest leads (V2 to V5) on ECG might be useful for predicting US in patients with ILRs.
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...