Remote interrogation and reprogramming of cardiac implantable electronic devices using a custom multivendor solution
Sylvain Ploux1, Marc Strik1, Fabrice Demonière2
1IHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, Pessac-Bordeaux, France; Bordeaux University Hospital (CHU), Cardio-Thoracic Unit, Pessac, France.
Background:
Until recently, remote interrogation and reprogramming of therapeutic cardiac implantable electronic devices (CIEDs) have been virtually nonexistent owing to technical challenges and safety concerns. It could be extremely useful, in particular in case of emergencies, when patients live far from CIED professionals or during enforced physical distancing.
Objective:
We investigated the feasibility and safety of a custom solution for remote interrogation and reprogramming of CIEDs from various manufacturers in various clinically relevant situations.
Methods:
Our solution consists of remote controlling CIED programmers through screen capture and remote cursor control. In this multicenter feasibility study, the primary outcome was technical feasibility (% of success) and safety (absence of complication) of interrogation and reprogramming when indicated in clinically driven encounters.
Results:
A total of 115 remote interrogations were performed in 110 patients: within the hospital (n = 73), medium range (50-100 km; n = 22), and long range (>5000 km; n = 20). Implanted devices were pacemakers (n = 85) and implantable cardioverter-defibrillators (n = 25) from Abbott (N=13), Biotronik (n = 65), and MicroPort (n = 32). Patients were located in the outpatient clinic, cardiology department, radiology department (magnetic resonance imaging), operating room (per implantation), and intensive care unit. Teleworking was performed in 39 cases. Complete CIED interrogations succeeded in all patients with reprogramming in 56 of 115 sessions (49%). No clinical or technical complications occurred. The time lag for screen interaction was below 1 second.
Conclusion:
Remote interrogation and reprogramming of CIEDs are feasible and safe across disparate clinical contexts and distances. This strategy may enhance health care access and facilitate medical training, tele-expertise, and telework worldwide.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
04:24A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Cardiomyopathy V: Interprofessional Care
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...