Related Experiment Videos
An evaluation of the Cell-Dyn 1000
P Flanagan1, A M Lane, A Delamare
1Department of Haematology, Westminster Hospital, London.
Clinical and Laboratory Haematology
|January 1, 1987
Abstract:
The Cell-Dyn 1000 is a new eight parameter fully automatic cell counter. This was evaluated following the ICSH guidelines. Scientific evaluation was generally satisfactory, largely conforming to the manufacturer's specifications. A number of potential microbiological safety hazards were noted; many could be overcome by minor modifications. Throughput and start-up/close-down times were acceptable. However, the reliability of the tested machine was felt to be less than satisfactory although this may be corrected by further minor machine modifications.