Major Complications and Failure Modes of the Angiosculpt Scoring Balloon Catheter: Analysis of the MAUDE Database

John S Dayco1, Heba Osman1, Talal Almas2

  • 1Wayne State University, Detroit Medical Center, MI.

Insights

The Angiosculpt scoring balloon catheter shows a low complication rate for treating calcified vascular lesions. Most issues involved device tip breaks or balloon ruptures, particularly in superficial femoral artery interventions.

Area of Science:

  • Cardiovascular Interventions
  • Medical Device Safety
  • Vascular Surgery

Background:

  • The Angiosculpt scoring balloon catheter is FDA-approved for treating calcified vascular lesions.
  • Understanding its complication profile is crucial for safe clinical application.

Purpose of the Study:

  • To characterize complication rates, failure modes, and outcomes of the Angiosculpt catheter.
  • To analyze device performance in severely calcified stenotic vascular lesions.

Main Methods:

  • Analysis of Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database events from October 2013 to December 2020.
  • Investigated 248 reported complications associated with the Angiosculpt catheter.

Main Results:

  • Most complications occurred in the superficial femoral artery (SFA) and left anterior descending artery.
  • Device damage (tip break, balloon rupture) and withdrawal difficulties were primary issues.
  • Balloon rupture rates were significantly higher in calcified vessels (60.6%) and SFA interventions (39.4%).

Conclusions:

  • The Angiosculpt catheter demonstrates a relatively low overall complication rate.
  • Complications are predominantly linked to device integrity and withdrawal challenges in calcified lesions, especially within the SFA.