[Translated article] Significant improvement in axial load stability with the pre-tensioned pelvic external fixator:
F J Durán Garrido1, A Pérez de la Blanca2, M Lombardo Torre1
1Unidad de Cirugía Ortopédica y Traumatología, Hospital Universitario Virgen de La Victoria, Málaga, Spain.
Objective:
The aim of our work is to check if the use of a pre-tensed bar connected to 2 supra-acetabular Schanz pins is more stable in a pelvic with a tile C1 injury.
Material And Methods:
We used synthetic pelvis (N = 5) with a unstable vertical injury, stabilized with both the standard and the pre-tensed system. They both were submitted to axial loading. We measured applied forces and the position of markers that were positioned in the sacroiliac and symphysis joints. We determined the relative movements between injured surfaces, rigidity, resistance and when each system failed.
Results:
When using the pre-tensed fixator, the rigidity was much higher in the sacro-iliac joint(P=.043) being 3.45 times higher, being also 2.06 times higher in pubic symphysis (P=.043). The resistance against failure was 2 times higher too (P=.043).
Discussion:
The increased axial stability with the pre-tensed external fixator accords with its ability to make active compression in both posterior and anterior elements of the pelvic ring, which was previously demonstrated. This result supports the preliminary clinic outcomes that were obtained.
Conclusions:
The pre-tensed external fixator increases the axial stability of the pelvic ring's posterior elements in a synthethic pelvis with a Tile C1 injury, without decreasing it's capacity to achieve an anterior stability.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Related Concept Videos
Eccentric Axial Loading in a Plane of Symmetry
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Normal Strain under Axial Loading