Background:
Defibrillators are medical devices (MDs) used in the most critical situations, hence their performance must be ensured at all times. This requires defibrillators to be subjected to regular performance assessments after they have been placed on the market. The new Medical Device Regulation (MDR) defines medical device post-market surveillance (PMS) as performed by independent, third-party, notified bodies more strategically in hope to improve traceability of device performance. However, there is still an apparent gap in terms of standardised conformity assessment testing methods.
Objective:
This paper proposes a novel evidence-based method for conformity assessment testing of defibrillators.
Method:
The proposed method is developed in accordance with the International Organisation of Legal Metrology (OIML) guidelines and is intended to be used for conformity assessment testing of defibrillators for post-market surveillance purposes.
Results:
The developed method was validated from 2018 to 2021 in healthcare institutions of all levels. The results obtained during validation suggest that conformity assessment testing of defibrillators as a method used during PMS contributes to significant improvement in devices' accuracy and reliability.
Conclusion:
A standardized approach in conformity assessment testing of defibrillators during PMS, besides increasing reliability of the devices, is the first step in the digital transformation of the management of these devices in healthcare institutions opening possibility for use of artificial intelligence.