Comparison of three methods for identifying nonfermenting gram-negative rods
Canadian Journal of Microbiology
|October 1, 1978
Summary
The Oxi-Ferm system demonstrated superior accuracy in identifying nonfermenting gram-negative rods compared to the API system, despite some biochemical discrepancies. Both systems struggled with specific bacterial species like Pseudomonas aeruginosa and P. cepacia.
Area of Science:
- Clinical Microbiology
- Bacterial Identification
- Diagnostic Systems
Background:
- Accurate identification of nonfermenting gram-negative rods is crucial for patient care.
- Conventional methods are labor-intensive, necessitating evaluation of automated systems.
- The API and Oxi-Ferm systems are widely used for bacterial identification.
Purpose of the Study:
- To compare the accuracy of the API and Oxi-Ferm systems against conventional methods for identifying nonfermenting gram-negative rods.
- To identify specific challenges encountered by each system in bacterial identification.
Main Methods:
- Seventy-six nonfermenting gram-negative rod strains were tested using the API system and conventional media.
- Fifty-one strains were additionally tested using the Oxi-Ferm system.
- Identification results from automated systems were compared to those from conventional methods.
Main Results:
- The Oxi-Ferm system achieved 72% agreement with conventional methods, while the API system showed only 41% agreement.
- Both systems exhibited difficulties in identifying Pseudomonas aeruginosa (API) and P. cepacia.
- Oxi-Ferm had more individual biochemical discrepancies but higher overall agreement.
Conclusions:
- The Oxi-Ferm system offers a more accurate identification of nonfermenting gram-negative rods compared to the API system.
- Further refinement of both systems may be needed to improve identification of challenging species.
- Automated systems provide valuable alternatives to conventional methods in clinical microbiology labs.