Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking
Published on: February 12, 2011
Effects of spatial and temporal resolution on cardiovascular magnetic resonance feature tracking measurements using a
David M Adams1, Redha Boubertakh2, Marc Eric Miquel1,3,4
1Clinical Physics, Barts Health NHS Trust, London, United Kingdom.
Objectives:
To develop a single-slice numerical phantom with known myocardial motion, at several temporal and in-plane spatial resolutions, for testing and comparison of Cardiovascular Magnetic Resonance (CMR) feature tracking (FT) software.
Methods:
The phantom was developed based on CMR acquisitions of one volunteer (acquired cine, tagging cine, T1 map, T2 map, proton density weighted image). The numerical MRI simulator JEMRIS was used, and the phantom was generated at several in-plane spatial resolutions (1.4 × 1.4 mm2 to 3.0 × 3.0 mm2) and temporal resolutions (20 to 40 cardiac phases). Two feature tracking software packages were tested: Medical Image Tracking Toolbox (MITT) and two versions of cvi42 (v5.3.8 and v5.13.7). The effect of resolution on strain results was investigated with reference to ground-truth radial and circumferential strain.
Results:
Peak radial strain was consistently undermeasured more for cvi42 v5.13.7 than for v5.3.8. Increased pixel size produced a trend of increased difference from ground-truth peak strain, with the largest changes for cvi42 obtained using v5.13.7 between 1.4 × 1.4 mm2 and 3.0 × 3.0 mm2, at 9.17 percentage points (radial) and 8.42 percentage points (circumferential).
Conclusions:
The results corroborate the presence of intervendor differences in feature tracking results and show the magnitude of strain differences between software versions.
Advances In Knowledge:
This study shows how temporal and in-plane spatial resolution can affect feature tracking with reference to the ground-truth strain of a numerical phantom. Results reaffirm the need for numerical phantom development for the validation and testing of FT software.