Background:
Cardiac implantable electronic device (CIED) infections have a high morbidity and mortality and are an indication of device extraction. As a replacement, leadless pacemakers (LPs) may be preferable due to a low infection risk, but mid-term data on reinfections is lacking. Moreover, early LP reimplantation in pacemaker-dependent patients would circumvent the need for temporary pacemakers.
Methods:
We included all patients with LP implantation as a replacement for an infected CIED, between January 2013 and December 2021. The occurrence of reinfection was assessed during standard follow-up visits.
Results:
Twenty-nine patients (mean age 81 ± 9 years) were included, of which 21 (73%) had a pocket infection, 7 (24%) endocarditis, and 1 (3%) a systemic infection without endocarditis. All LP implantations were successful. LPs were implanted before extraction (n = 4, 13%), simultaneously with extraction (n = 5, 17%) and after extraction (n = 20, 70%). No reinfection occurred during the follow-up of median 32 months (IQR 13-66 months). Repeat blood cultures obtained in 9 (30%) patients and transthoracic echocardiography in all 7 patients with pacemaker endocarditis were negative for reinfection. In a subset of 6 LPs extracted during follow-up due to early battery depletion, prophylactically after the battery advisory or due to non-capture (median 36 months (range 0-67 months) post-implantation), histopathologic examination of tissues around the LPs showed no signs of infection.
Conclusions:
After replacing infected CIEDs for an LP, no reinfections occurred in over 2.5 years follow-up. These results confirm that in case of CIED infection, the LP is an appealing replacement device. LP implantation before CIED extraction is feasible.