Related Experiment Video
Updated: Aug 16, 2025

Evaluation of Commercial-Off-The-Shelf Wrist Wearables to Estimate Stress on Students
Published on: June 16, 2018
Comparing a Fitbit Wearable to an Electrocardiogram Gold Standard as a Measure of Heart Rate Under Psychological
Joel Gagnon1, Michelle Khau2, Léandre Lavoie-Hudon2
1School of Psychology, Faculty of Social Sciences, Laval University, Québec, QC, Canada.
Background:
Wearable devices collect physiological and behavioral data that have the potential to identify individuals at risk of declining mental health and well-being. Past research has mainly focused on assessing the accuracy and the agreement of heart rate (HR) measurement of wearables under different physical exercise conditions. However, the capacity of wearables to sense physiological changes, assessed by increasing HR, caused by a stressful event has not been thoroughly studied.
Objective:
This study followed 3 objectives: (1) to test the ability of a wearable device (Fitbit Versa 2) to sense an increase in HR upon induction of psychological stress in the laboratory; (2) to assess the accuracy of the wearable device to capture short-term HR variations caused by psychological stress compared to a gold-standard electrocardiogram (ECG) measure (Biopac); and (3) to quantify the degree of agreement between the wearable device and the gold-standard ECG measure across different experimental conditions.
Methods:
Participants underwent the Trier Social Stress Test protocol, which consists of an oral phase, an arithmetic stress phase, an anticipation phase, and 2 relaxation phases (at the beginning and the end). During the stress protocol, the participants wore a Fitbit Versa 2 and were also connected to a Biopac. A mixed-effect modeling approach was used (1) to assess the effect of experimental conditions on HR, (2) to estimate several metrics of accuracy, and (3) to assess the agreement: the Bland-Altman limits of agreement (LoA), the concordance correlation coefficient, the coverage probability, the total deviation index, and the coefficient of an individual agreement. Mean absolute error and mean absolute percent error were calculated as accuracy indices.
Results:
A total of 34 university students were recruited for this study (64% of participants were female with a mean age of 26.8 years, SD 8.3). Overall, the results showed significant HR variations across experimental phases. Post hoc tests revealed significant pairwise differences for all phases. Accuracy analyses revealed acceptable accuracy according to the analyzed metrics of accuracy for the Fitbit Versa 2 to capture the short-term variations in psychological stress levels. However, poor indices of agreement between the Fitbit Versa 2 and the Biopac were found.
Conclusions:
Overall, the results support the use of the Fitbit Versa 2 to capture short-term stress variations. The Fitbit device showed acceptable levels of accuracy but poor agreement with an ECG gold standard. Greater inaccuracy and smaller agreement were found for stressful experimental conditions that induced a higher HR. Fitbit devices can be used in research to measure HR variations caused by stress, although they cannot replace an ECG instrument when precision is of utmost importance.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
10:45A Community-based Stress Management Program: Using Wearable Devices to Assess Whole Body Physiological Responses in Non-laboratory Settings
Published on: January 22, 2018
Related Concept Videos
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...