Related Experiment Video
Updated: Aug 16, 2025

Designing CAD/CAM Surgical Guides for Maxillary Reconstruction Using an In-house Approach
Published on: August 24, 2018
Virtual planning and 3D-printed guides for mandibular reconstruction: Factors impacting accuracy
Donald J Annino1,2, Rosh K Sethi1,2,3, Elisabeth E Hansen4
1Division of Otolaryngology Brigham and Women's Hospital, Dana-Farber Cancer Institute Boston Massachusetts USA.
Objectives:
Examine accuracy and factors impacting accuracy for mandibular reconstruction with virtual surgical planning, 3D printed osteotomy guides and preoperatively bent mandibular reconstruction plate (VSP/3Dprinted-guide/plate).
Method:
Retrospective review of osseous-free-flap mandibular reconstructions with VSP/3Dprinted-guide/plate between January 2015 and July 2020 at a single academic medical center.Patient demographics, disease, and treatment variables were extracted. Accuracy was assessed by 3D-model-overlay with cephalometric and donor-bone segment length measurements. Multivariate analyses were performed to determine factors impacting cephalometric accuracy.
Results:
60 cases met criteria: 41 (68%) cancer, 14 (23%) osteoradionecrosis (ORN), 5 (8%) secondary mandibular reconstruction. Thirteen cases (22%) were Brown class III or IV. Thirty-nine cases (65%) had ≥2 flap bone segments. Average donor-bone length was 82 mm (SD: 28). 3D-model-overlay accuracy demonstrated minimal deviation between planned and actual reconstruction: intercondylar distance = 2.10 mm (SD: 2.2); intergonial distance = 2.23 mm (SD: 1.9); anterior-posterior distance (APD) = 1.76 mm (SD: 1.5); gonial angle (GA) = 3.11 degrees (SD: 2.4). Mean change in donor-bone segment length inferiorly was 2.67 mm (SD: 2.6) and superiorly 3.27 mm (SD: 3.2). Higher number of donor-bone segments was associated with decreased accuracy in GA (p = .023) and longer donor-bone length was associated with decreased accuracy in APD (p = .031).
Conclusion:
To our knowledge this is the largest series assessing surgical accuracy of VSP/3Dprinted-guide/plate for osseous-free-flap mandibular reconstruction. We demonstrate highly accurate results, with increased number of donor-bone segments and donor-bone length associated with decreased accuracy. Our findings further support VSP/3Dprinted-guide/plate as a reliable and accurate tool for mandibular reconstruction.
Level Of Evidence:
Level 4.