A Feasibility Study for Immediate Histological Assessment of Various Skin Biopsies Using Ex Vivo Confocal Laser
Hanna Ogrzewalla1, Matthias Möhrle2, Gisela Metzler3
1Department of Dermatology, Eberhardt Karls Universität, 72074 Tübingen, Germany.
Diagnostics (Basel, Switzerland)
|December 23, 2022
Summary
Confocal laser scanning microscopy (CLSM) shows potential for digital dermatologic diagnosis but has limitations. Diagnostic accuracy varies across tumor types, suggesting it is not yet suitable for all applications.
Area of Science:
- Dermatopathology
- Digital Microscopy
- Diagnostic Imaging
Background:
- Confocal laser scanning microscopy (CLSM) offers a digital alternative to traditional formalin-fixed, paraffin-embedded (FFPE) and hematoxylin-eosin (H&E) stained slides.
- This study investigates the diagnostic accuracy of digitally stained CLSM scans against H&E slides in real-world dermatologic conditions.
Purpose of the Study:
- To evaluate the diagnostic accuracy of digitally stained CLSM scans.
- To compare CLSM diagnostic performance with conventional H&E staining for various dermatologic diseases.
Main Methods:
- Dermatologic patient samples were scanned using CLSM.
- Diagnoses were confirmed using FFPE and H&E slides by two dermatopathologists.
- Sensitivity and specificity of CLSM diagnoses were calculated for three distinct diagnostic groups.
Main Results:
- Epithelial tumors (n=132): Sensitivity 64.3%/83.9%, Specificity 84.2%/71.1%.
- Melanocytic tumors (n=86): Sensitivity 19.1%/85.1%, Specificity 96.3%/66.7%.
- Other tumors/cysts and inflammatory dermatoses (n=42): Sensitivity 96.4%/96.8%, Specificity 57.1%/45.5%.
Conclusions:
- Digitally stained CLSM scans demonstrate both capabilities and constraints for widespread dermatologic use.
- Variable diagnostic accuracy across different conditions indicates CLSM is not universally applicable at this time.