A Retrospective, Observational Study of Catheter-Associated Urinary Tract Infection Events Post-Implementation of a
Patrick Brockway1, David M Hill1, Vanessa Moll2,3
1Firefighters Burn Center, Regional One Health, Memphis, TN 38103, USA.
Abstract:
Objective: A quality improvement study to assess catheter-associated urinary tract infection (CAUTI) rate post-implementation of a bladder catheter with integrated active drain line urine clearance and automated intra-abdominal pressure monitoring in a burn intensive care unit (ICU). DESIGN: Eight-year retrospective before and after study (2015−2022). Setting: A single American Burn Association-verified Burn Center with 14 inpatient beds. Patients: Patients meeting criteria for admission to a Burn Center. Methods: Retrospective cohort study following the implementation of a novel urine output monitoring system with integrated drain line and urine clearance. Data from a 48-month (from January 2015−December 2018) historical control (period 1) were compared to data from a 28-month (from January 2020 to April 2022) post-implementation period (period 2). Pre- and post-implementation CAUTI event incidences were compared. Patients were transferred from outside hospitals with gravity bladder. A distinction in the chart between catheter types was impossible. Charts were reviewed to characterize patients with CAUTI events. Results: A total of 42 CAUTIs in 2243 patients were identified using the National Health and Safety Network (NHSN) definition during the analyzed period. There were 40 CAUTI events in period 1 and two CAUTIs in period 2. The incidence of CAUTI events pre-implementation was 0.030 (mean of 10 CAUTI events per year) compared to 0.002 (mean of 1 CAUTI event per year) post-implementation of an automatic drain line clearing UO monitoring system showing a significant reduction in CAUTI events (p < 0.01, risk ratio novel vs. gravity bladder catheter 0.071, 95% confidence interval: 0.017−0.294). Conclusions: CAUTIs were reduced in the period following the implementation of a novel urinary catheter system with an integrated active drain line and urine clearance in burn patients.
Related Concept Videos
Urinary Tract Infection III: Diagnostic Studies and Interprofessional Care
Urine Studies II: Urine Culture and Sensitivity Test
Urinary Tract Infection II: Pathophysiology
Urinary Tract Infection IV: Nursing Management
Urinary Tract Infection I: Introduction
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...