Using Micro-Electro-Mechanical Systems MEMS to Develop Diagnostic Tools
Published on: October 1, 2007
Design and Fabrication of an Integrated Hollow Concave Cilium MEMS Cardiac Sound Sensor
Bo Wang1, Pengcheng Shi1, Yuhua Yang1
1State Key Laboratory of Dynamic Measurement Technology, School of Instrument and Electronics, North University of China, Taiyuan 030051, China.
Abstract:
In light of a need for low-frequency, high sensitivity and broadband cardiac murmur signal detection, the present work puts forward an integrated MEMS-based heart sound sensor with a hollow concave ciliary micro-structure. The advantages of a hollow MEMS structure, in contrast to planar ciliated micro-structures, are that it reduces the ciliated mass and enhances the operating bandwidth. Meanwhile, the area of acoustic-wave reception is enlarged by the concave architecture, thereby enhancing the sensitivity at low frequencies. By rationally designing the acoustic encapsulation, the loss of heart acoustic distortion and weak cardiac murmurs is reduced. As demonstrated by experimentation, the proposed hollow MEMS structure cardiac sound sensor has a sensitivity of up to -206.9 dB at 200 Hz, showing 6.5 dB and 170 Hz increases in the sensitivity and operating bandwidth, respectively, in contrast to the planar ciliated MEMS sensor. The SNR of the sensor is 26.471 dB, showing good detectability for cardiac sounds.