Construct Validity and Reproducibility of Handheld Ultrasound Devices in Carotid Artery Diameter Measurement
Loes Willems1, Jenske Vermeulen2, Anouk Wiegerinck3
1Department of Surgery, Radboud University Medical Centre, Nijmegen, The Netherlands.
Abstract:
The construct validity and reproducibility of three commonly used handheld ultrasound (US) devices in measuring carotid arterial diameter was evaluated: Telemed MicrUs EXT-1H (Telemed, Vilnius, Lithuania), Butterfly iQ (Butterfly Network, Inc., Guilford, CT, USA) and Philips Lumify (Philips Healthcare, Best, The Netherlands). An in vitro setup was built to evaluate construct validity, compared with high-end US, and intra-observer variability of handheld US devices. Handheld devices showed a mean difference of 0.023 ± 0.030 cm, 0.012 ± 0.037 cm and 0.009 ± 0.046 cm for, respectively, Telemed, Butterfly and Lumify in comparison with high-end US devices. Intraclass agreement with the high-end system as well as intra-observer variability for handheld US devices was classified as excellent, with all values greater than 0.95. Subsequently, inter-observer variability of handheld US devices was investigated in an in vivo setup with 20 healthy volunteers. Inter-observer variability was classified as excellent for Telemed (0.901), good for Lumify (0.827) and moderate for Butterfly (0.684) with a difference of, respectively, 0.005 ± 0.031 cm, 0.020 ± 0.050 cm and -0.003 ± 0.033 cm. In conclusion, handheld US devices demonstrated an excellent construct validity and intra-observer variability. Additionally, excellent-to-good inter-observer variability for Telemed and Lumify was observed, and Butterfly demonstrated a moderate inter-observer agreement. These results indicate that handheld US devices are effective for measuring carotid arterial diameter.
More Related Videos
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement