Multivendor Evaluation of Automated MRI Postprocessing of Biventricular Size and Function for Children With and
Jelle P G van der Ven1,2, Wouter van Genuchten1,2, Zaheda Sadighy1
1Department of Pediatrics, Division of Cardiology, Erasmus University Medical Center-Sophia Children's Hospital, Rotterdam, the Netherlands.
Background:
Manually segmenting cardiac structures is time-consuming and produces variability in MRI assessments. Automated segmentation could solve this. However, current software is developed for adults without congenital heart defects (CHD).
Purpose:
To evaluate automated segmentation of left ventricle (LV) and right ventricle (RV) for pediatric MRI studies.
Study Type:
Retrospective comparative study.
Population:
Twenty children per group of: healthy children, LV-CHD, tetralogy of Fallot (ToF), and univentricular CHD, aged 11.7 [8.9-16.0], 14.2 [10.6-15.7], 14.6 [11.6-16.4], and 12.2 [10.2-14.9] years, respectively.
Sequence/Field Strength:
Balanced steady-state free precession at 1.5 T.
Assessment:
Biventricular volumes and masses were calculated from a short-axis stack of images, which were segmented manually and using two fully automated software suites (Medis Suite 3.2, Medis, Leiden, the Netherlands and SuiteHeart 5.0, Neosoft LLC, Pewaukee, USA). Fully automated segmentations were manually adjusted to provide two further sets of segmentations. Fully automated and adjusted automated segmentation were compared to manual segmentation. Segmentation times and reproducibility for each method were assessed.
Statistical Tests:
Bland Altman analysis and intraclass correlation coefficients (ICC) were used to compare volumes and masses between methods. Postprocessing times were compared by paired t-tests.
Results:
Fully automated methods provided good segmentation (ICC > 0.90 compared to manual segmentation) for the LV in the healthy and left-sided CHD groups (eg LV-EDV difference for healthy children 1.4 ± 11.5 mL, ICC: 0.97, for Medis and 3.0 ± 12.2 mL, ICC: 0.96 for SuiteHeart). Both automated methods gave larger errors (ICC: 0.62-0.94) for the RV in these populations, and for all structures in the ToF and univentricular CHD groups. Adjusted automated segmentation agreed well with manual segmentation (ICC: 0.71-1.00), improved reproducibility and reduced segmentation time in all patient groups, compared to manual segmentation.
Data Conclusion:
Fully automated segmentation eliminates observer variability but may produce large errors compared to manual segmentation. Manual adjustments reduce these errors, improve reproducibility, and reduce postprocessing times compared to manual segmentation. Adjusted automated segmentation is reasonable in children with and without CHD.
Evidence Level:
3.
Technical Efficacy:
Stage 2.
More Related Videos
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
08:21Biventricular Assessment of Cardiac Function and Pressure-Volume Loops by Closed-Chest Catheterization in Mice
Published on: June 15, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...