Some patients with fixed-rate demand pacemakers may benefit from conversion to a rate-responsive pacing system. In this report, we describe a technique by which the effect of rate-responsive pacing can be assessed before replacing the pacemaker system. An externally positioned activity-mode physiologic pulse generator was used to drive a previously implanted ventricular demand pacemaker programmed to the triggered mode. Comparison of exercise capacity under both conditions suggested a substantial benefit from rate-responsiveness.