Evaluation of a workflow for cone-beam CT-guided online adaptive palliative radiotherapy planned using diagnostic CT
Koen J Nelissen1,2, Eva Versteijne1,2, Suresh Senan1,2
1Department of Radiation Oncology, Amsterdam UMC location Vrije Universiteit Amsterdam, Amsterdam, The Netherlands.
Purpose:
Single-visit radiotherapy (RT) is beneficial for patients requiring pain control and can limit interruptions to systemic treatments. However, the requirement for a dedicated planning CT (pCT)-scan can result in treatment delays. We developed a workflow involving preplanning on available diagnostic CT (dCT) imaging, followed by online plan adaption using a cone-beam CT (CBCT)-scan prior to RT-delivery, in order to account for any changes in anatomy and target position.
Methods:
Patients previously treated with palliative RT for bone metastases were selected from our hospital database. Patient dCT-images were deformed to treatment CBCTs in the Ethos platform (Varian Medical Systems) and a synthetic CT (sCT) generated. Treatment quality was analyzed by comparing a coverage of the V95% of the planning/clinical target volume and different organ-at-risk (OAR) doses between adapted and initial clinical treatment plans. Doses were recalculated on the CBCT and sCT in a separate treatment planning system. Adapted plan doses were measured on-couch using an anthropomorphic phantom with a Gafchromic EBT3 dosimetric film and compared to dose calculations.
Results:
All adapted treatment plans met the clinical goals for target and OARs and outperformed the original treatment plans calculated on the (daily) sCT. Differences in V95% of the target volume coverage between the initial and adapted treatments were <0.2%. Dose recalculations on CBCT and sCT were comparable, and the average gamma pass rate (3%/2 mm) of dosimetric measurements was 98.8%.
Conclusions:
Online daily adaptive RT using dCTs instead of a dedicated pCT is feasible using the Ethos platform. This workflow has now been implemented clinically.