Lung Cancer Screening Using Clinical Photon-Counting Detector Computed Tomography and Energy-Integrating-Detector
Akitoshi Inoue1, Tucker F Johnson, Lara A Walkoff
1From the Department of Radiology, Mayo Clinic, Rochester, MN.
Objective:
To evaluate the diagnostic quality of photon-counting detector (PCD) computed tomography (CT) in patients undergoing lung cancer screening compared with conventional energy-integrating detector (EID) CT in a prospective multireader study.
Materials:
Patients undergoing lung cancer screening with conventional EID-CT were prospectively enrolled and scanned on a PCD-CT system using similar automatic exposure control settings and reconstruction kernels. Three thoracic radiologists blinded to CT system compared PCD-CT and EID-CT images and scored examinations using a 5-point Likert comparison score (-2 [left image is worse] to +2 [left image is better]) for artifacts, sharpness, image noise, diagnostic image quality, emphysema visualization, and lung nodule evaluation focusing on the border. Post hoc correction of Likert scores was performed such that they reflected PCD-CT performance in comparison to EID-CT. A nonreader radiologist measured objective image noise.
Results:
Thirty-three patients (mean, 66.9 ± 5.6 years; 11 female; body mass index; 30.1 ± 5.1 kg/m 2 ) were enrolled. Mean volume CT dose index for PCD-CT was lower (0.61 ± 0.21 vs 0.73 ± 0.22; P < 0.001). Pooled reader results showed significant differences between imaging modalities for all comparative rankings ( P < 0.001), with PCD-CT favored for sharpness, image noise, image quality, and emphysema visualization and lung nodule border, but not artifacts. Photon-counting detector CT had significantly lower image noise (74.4 ± 10.5 HU vs 80.1 ± 8.6 HU; P = 0.048).
Conclusions:
Photon-counting detector CT with similar acquisition and reconstruction settings demonstrated improved image quality and less noise despite lower radiation dose, with improved ability to depict pulmonary emphysema and lung nodule borders compared with EID-CT at low-dose lung cancer CT screening.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation I: X-ray and CT