Reliability and agreement of apparent chord mu measurements between static and dynamic evaluations
Elena Martínez-Plaza1, Mario Marcos, Alberto López-de la Rosa
1From the Instituto de Oftalmobiología Aplicada (IOBA), Universidad de Valladolid, Valladolid, Spain (Martínez-Plaza, Marcos, López-de la Rosa, López-Miguel, Maldonado); Red Temática de Investigación Colaborativa en Oftalmología (OftaRed), Instituto de Salud Carlos III, Madrid, Spain (Martínez-Plaza, López-Miguel, Maldonado).
Purpose:
To assess the repeatability and agreement of Cartesian coordinates and the length of apparent chord mu and pupil diameter measurements during static (Galilei G4) and dynamic (Topolyzer Vario) evaluations.
Setting:
IOBA-Eye Institute, Valladolid, Spain.
Design:
Case series.
Methods:
3 consecutive measurements per scenario (Galilei G4 and Topolyzer Vario under low mesopic and photopic conditions) were performed by the same clinician. The intrasession repeatability was assessed using the within-subject SD (Sw), the precision, the coefficient of variation, and the intraclass correlation coefficient (ICC). The agreement was analyzed using repeated-measures analysis of variance and the Bland-Altman method.
Results:
Thirty-seven healthy participants were recruited. The Sw values for chord mu parameters and pupil diameter ranged from 0.01 to 0.03 and 0.08 to 0.21, respectively. The ICC was ≥0.89 for all parameters. Galilei G4 and Topolyzer Vario under low mesopic and photopic conditions provided significantly different measures of apparent chord mu length (0.23 ± 0.11 mm, 0.30 ± 0.10 mm, and 0.25 ± 0.11 mm, respectively, P ≤ .02), X-coordinate (-0.18 ± 0.12 mm, -0.27 ± 0.11 mm, and -0.21 ± 0.12 mm, respectively, P < .001), and pupil diameter (3.38 ± 0.50 mm, 6.29 ± 0.60 mm, and 3.04 ± 0.41 mm, respectively, P < .001). Y-coordinate values obtained by Galilei G4 and Topolyzer Vario under low mesopic conditions were significantly different (0.06 ± 0.13 mm vs 0.03 ± 0.11 mm, respectively, P = .02), in contrast to Galilei G4 and Topolyzer Vario under photopic conditions (0.05 ± 0.13 mm, P = .82) and both illumination conditions of Topolyzer Vario (P ≥ .23).
Conclusions:
Galilei G4 and Topolyzer Vario provide consistent measurements of apparent chord mu Cartesian coordinates and length, as well as pupil diameter; however, the measurements are not interchangeable. Ophthalmic surgeons should consider these findings when planning customized intraocular lens implantation and refractive surgery procedures.
Related Concept Videos
Dynamic Modulus of Elasticity of Concrete
The sonic test is a common method to determine the dynamic modulus. In this test, a concrete beam, sized either 6 x 6 x 30 inches or 4 x 4 x 20 inches, is clamped at its center. Vibrations are initiated at one end of the beam by an electromagnetic exciter unit powered by...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: