Related Experiment Video
Updated: Aug 16, 2025

Rapid Point-of-Care Assay of Enoxaparin Anticoagulant Efficacy in Whole Blood
Published on: October 12, 2012
Resolving DOAC interference on antithrombin activity testing on a FXa based method by the use of activated carbon
Jill Vercruyssen1, Peter Meeus1, Els Bailleul1
1Laboratory of Biochemistry and Hematology, Onze-Lieve-Vrouw Ziekenhuis, Moorselbaan 164, 9300 Aalst, Belgium.
Activated carbon (AC) effectively removes interference from direct oral anticoagulants (DOACs) in antithrombin (AT) activity assays. This ensures accurate measurement of AT levels, even in the presence of these common anticoagulant drugs.
Area of Science:
- Clinical Chemistry
- Hematology
- Pharmacology
Background:
- Direct oral anticoagulants (DOACs) can interfere with antithrombin (AT) activity assays, leading to inaccurate results.
- Activated carbon (AC) has demonstrated efficacy in removing DOAC interference in other coagulation assays.
- The utility of AC in resolving DOAC interference specifically in AT assays requires investigation.
Purpose of the Study:
- To evaluate the effectiveness of activated carbon (AC) in removing direct oral anticoagulant (DOAC) interference in antithrombin (AT) activity measurements.
- To determine if AC can restore accurate AT activity readings in the presence of various DOACs.
Main Methods:
- Normal pooled plasma (NPP) was diluted to achieve AT activity levels of 25%, 50%, and 75%.
- Diluted NPP samples were spiked with different concentrations (100, 250, 500 ng/ml) of four DOACs (apixaban, edoxaban, dabigatran, rivaroxaban).
- AT activity was measured using a FXa-based assay before and after treatment with 20 mg/ml AC.
Main Results:
- All tested DOAC concentrations were below the limit of quantification after AC treatment.
- Activated carbon successfully removed DOAC interference, enabling accurate diagnosis of low AT levels across all dilutions.
- The degree of DOAC interference was directly proportional to the DOAC concentration, with anti-FXa DOACs showing the most significant impact on the assay.
Conclusions:
- Activated carbon is an effective method for eliminating interference from anti-FXa direct oral anticoagulants in FXa-based antithrombin activity assays.
- AC treatment allows for reliable assessment of antithrombin activity in patients treated with specific DOACs.
More Related Videos
12:55Quantification of Humic and Fulvic Acids in Humate Ores, DOC, Humified Materials and Humic Substance-Containing Commercial Products
Published on: March 18, 2022
11:56A Rapid and Specific Microplate Assay for the Determination of Intra- and Extracellular Ascorbate in Cultured Cells
Published on: April 11, 2014
Related Concept Videos
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Venous Thrombosis III: Interprofessional Care
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Masking and Demasking Agents
There are many masking agents, such as cyanide, fluoride, triethanolamine, thiourea, and 2,3-bis(sulfanyl)propan-1-ol (formerly 2,3-dimercapto-1-propanol), with the masking agent chosen based on...