Objectives:
The aim of this study was to fractographically examine clinically fractured fixation screws used to fix implant supported restorations and to estimate stress at failure.
Methods:
20 clinically fractured titanium implant fixation screws were collected and analyzed under scanning electron microscope in order to locate critical crack origin and calculate stress at fracture. Principles of fractographic analysis were applied using relative equations and cause and conditions of screw fracture were calculated. X-rays and patients' records were collected after approval of necessary consent forms.
Results:
Fractographic analysis revealed location and dimensions of critical crack size for every fractured screw examined. Two patterns of screw fracture were identified; the first was related to improper seating of fixation screw leading to damage at screw threads and eventual fracture of screw body. The second was related to slow crack growth propagation causing final fracture of screw body. The calculated stress at failure ranged from 755 to 804 MPa. Patients' records revealed that 75% of fractured screws were related to single unit screw retained restorations.
Significance:
Single unit screw retained restorations and improperly seated implant abutments impose high stress concentration on fixation screw which may result in fixation screw fracture.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...