Assay validation for vilanterol and its metabolites in human urine with application for doping control analysis
Tina Panchal1, Sandra Baldwin2, Martin Østergaard3
1Bioanalysis, Immunogenicity and Biomarkers, GSK, Research and Development, Stevenage, UK.
Abstract:
A bioanalytical method for detecting the ultra-long-acting beta2 -agonist (U-LABA) inhaled vilanterol and its metabolites, GSK932009 and GW630200, in urine was developed to potentially monitor permitted therapeutic versus prohibited supratherapeutic use in sport. The World Anti-Doping Agency (WADA) has established urinary concentration thresholds for the beta2 -agonists salbutamol and formoterol. Therapeutic use of vilanterol (25 μg once daily) was recently permitted by WADA; however, there is no established decision limit for adverse analytical findings due to insufficient urine concentration data. In this study, we validated an assay to detect vilanterol in urine collected from four healthy male and female athletes 0-72 h who received inhaled corticosteroid fluticasone furoate/U-LABA vilanterol (800/100 μg) combination, four times the normal therapeutic dose. After administration, subjects performed 1 h of bike ergometer exercise. The experiment was conducted again after repeat dosing for 1 week. Our method utilised liquid chromatography with tandem mass spectrometry and was validated over urine concentrations of 5-5000 (vilanterol) and 50-50,000 pg/ml (GSK932009 and GW630200). Plasma samples were analysed for vilanterol, using a previously validated assay. The peak concentration values for urine vilanterol, GSK932009 and GW630200 were 9.5, 10.4 and 0.17 ng/ml, for single dosing, and 18.6, 19.5 and 0.20 ng/ml, for repeat dosing. Urine samples from four volunteers using the final validated method are reported, demonstrating this assay has sensitivity to detect vilanterol or GSK932009 in urine for ≥72 h post single or repeat dosing with 800/100 μg fluticasone furoate/vilanterol, whereas GW630200 was quantifiable ≤4 h post dose.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
06:14Optimized LC-MS/MS Method for the High-throughput Analysis of Clinical Samples of Ivacaftor, Its Major Metabolites, and Lumacaftor in Biological Fluids of Cystic Fibrosis Patients
Published on: October 15, 2017
Related Concept Videos
Urine Studies I: Urinalysis
Data Validation
Key parameters for method validation include: