Related Experiment Video
Updated: Aug 15, 2025

Surgical Training for the Implantation of Neocortical Microelectrode Arrays Using a Formaldehyde-fixed Human Cadaver Model
Published on: November 19, 2017
Use of a Pericranial Flap Technique for Deep Brain Stimulation Hardware Protection and Improved Cosmesis
Viktoras Palys1, Matthew Moser2, Shripadh Chitta3
1Department of Neurosurgery, University of Arkansas for Medical Sciences, Little Rock, AR, USA.
Objectives:
Deep brain stimulation (DBS) has become an established neuromodulation therapy; however, surgical site complications such as hardware skin erosion remain an important risk and can predispose to infection, requiring explantation of the system. Nuances of surgical technique can affect wound healing, cosmetic outcome, comfort, and risk of infection. In this study, we describe our experience with a layered closure technique using a vascularized pericranial flap for improving cosmesis and protection of the implanted hardware against skin erosion and infection.
Materials And Methods:
We retrospectively reviewed 636 individuals (746 lead implantations) who underwent DBS surgery by a single academic neurosurgeon between 2001 and 2020. A layered pericranial flap closure technique for the burr-hole and connector sites was instituted in 2015. We assessed the effects of a multimodal infection prevention approach that included the pericranial flap on hardware complication rates compared with the premultimodality cohort, and we report the nuances of the technique.
Results:
In our institutional experience, we found that implementation of a pericranial flap closure technique can enhance the subjective cosmetic result at the burr-hole cover site and increase patient comfort and satisfaction. In addition, we found a decrease in hardware infection rates in the current cohort with a multimodal infection prevention regimen that includes the pericranial-flap technique (n = 256, 2015-2020 period) to 1.2% (p = 0.006), from 6.9% in the earlier cohort (n = 490, 2001-2015 period).
Conclusions:
The report highlights the potential of a pericranial-flap closure technique as a surgical adjunct to improve DBS surgical site healing and cosmesis and may, as part of a multimodal strategy, contribute to decreased risk of skin breakdown and hardware infection.
More Related Videos
08:30Subdural Soft Electrocorticography ECoG Array Implantation and Long-Term Cortical Recording in Minipigs
Published on: March 31, 2023
08:42The DREAM Implant: A Lightweight, Modular, and Cost-Effective Implant System for Chronic Electrophysiology in Head-Fixed and Freely Behaving Mice
Published on: July 26, 2024