Related Experiment Video
Updated: Aug 15, 2025

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Re-Evaluation of Patient-Sourced Radiation Doses in PET/CT
Ahmet Murat Şenışık1, Handan Tanyıldızı Kökkülünk1, Mahmut Yüksel2
1Vocational School of Health Sciences, Department of Radiotherapy, Altınbaş University, Bakirkoy, Istanbul, Turkey.
Background:
New generation PET/CT devices provide quality images using low radiopharmaceutical activities. Dose monitoring is carried out for nuclear medicine personnel, other health personnel, and companions by determining the radiation dose emitted from low-activity patients to the environment. In particular, it is necessary to revise the working conditions of the personnel according to the radiation dose exposed.
Aim:
It was aimed to reevaluate the radiation dose rate transmitted to the environment from patients injected with 18F-FDG.
Materials And Methods:
A total of 31 patients (14F, 17M) who underwent 18F-FDG PET/CT imaging were included. The mean 18F-FDG activity of 7.26 ± 1.29 mCi was used for injection. After injection, radiation dose rates (mR/h) were measured at distances of 25, 50, 100, 150, and 200cm for 3 different periods from the level of the head, thorax, abdomen, and pelvis by using a GM counter. Additionally, biological samples such as urine and sweat were taken during 3 different periods. The activity amounts (μCi) in the samples were measured with a well-type counter.
Results:
Strong correlations were calculated between normalized dose rates obtained by all regions and time. Considering the nuclear medicine staff handling time with a PET/CT patient, the average dose received by staff was calculated between a range of 0.002-0.004 mSv/pt. The radiation dose exposed to the porter and nurse was calculated as 0.049 mSv/pt for the 2nd hour and 0.001-0.007 mSv/pt for the 4th hour, respectively. The companion was exposed to a dose between 0.073-0.147 mSv and 0.024-0.048 mSv for public transport and private car transportation after 4-6 hours of injection (for 30-60 min of travel duration), respectively. For inpatients, the received dose for porters, serving 20min from a distance of 30cm for the 2nd and 4th hours after the PET/CT scan, was 0.049 mSv/pt and 0.048 mSv/pt, respectively. And for nurses serving from a 50cm distance between 1-5 minutes, these values were found to be 0.001-0.007mSv/pt, 0.001-0.007mSv/pt, and 0.001-0.006mSv/pt, respectively.
Conclusion:
The radiation dose of nuclear medicine staff, porters, nurses, and companions are found to be below the recommended dose limit by the ICRP. According to our results, there is no need for any restrictions for patients, companions, or healthcare personnel in PET/CT units.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT