Infection of a Retained Peripheral Nerve Stimulation Lead: A Case Report

John J Finneran1, Bryan T Leek2, Brian M Ilfeld1

  • 1From the Departments of Anesthesiology.

A&A Practice
|January 4, 2023
PubMed

Insights

Infection of percutaneous peripheral nerve stimulation leads is rare. This case highlights a rare infection and lead fracture during removal, emphasizing prevention strategies.

Area of Science:

  • Pain Management
  • Neuromodulation
  • Medical Device Technology

Background:

  • Percutaneous peripheral nerve stimulation (PNS) offers prolonged analgesia.
  • Helically coiled leads have an exceptionally low infection rate, under 1 per 32,000 days.

Observation:

  • A rare case of infection involving a helically coiled PNS lead is reported.
  • Lead extraction required sedation and was complicated by intraoperative fracture, leaving a remnant.

Findings:

  • The infection was successfully managed with oral antibiotics.
  • Lead fracture during removal is a potential complication.

Implications:

  • This case underscores the importance of vigilance for rare complications in PNS.
  • Preventive measures, such as avoiding sling-to-lead insertion site overlap, are proposed.