Related Experiment Video
Updated: Aug 15, 2025

Author Spotlight: Advancing Cardiac Procedure Testing Prior to Embarking on Large Animal Studies
Published on: August 25, 2023
First-in-Human Dedicated Leaflet Splitting Device for Prevention of Coronary Obstruction in Transcatheter Aortic
Danny Dvir1, Martin B Leon2, Mohamed Abdel-Wahab3
1Department of Cardiology, Shaare Zedek Medical Center and Faculty of Medicine, Hebrew University of Jerusalem, Israel.
Background:
Coronary artery obstruction is a life-threatening complication of transcatheter aortic valve replacement (TAVR) procedures. Current preventive strategies are suboptimal.
Objectives:
The aim of this study was to describe bench testing and clinical experience with a novel device that splits valve leaflets that are at risk for causing coronary obstruction after TAVR, allowing normal coronary flow.
Methods:
The ShortCut device was initially tested in vitro and preclinically in a porcine model for functionality and safety. The device was subsequently offered to patients at elevated risk for coronary obstruction. Risk for coronary obstruction was based on computed tomography-based anatomical characteristics. Procedure success was determined as patient survival at 30 days with a functioning new valve, without stroke or coronary obstruction.
Results:
Following a successful completion of bench testing and preclinical trial, the device was used in 8 patients with failed bioprosthetic valves (median age 81 years; IQR: 72-85 years; 37.5% man) at 2 medical centers. A total of 11 leaflets were split: 5 patients (63.5%) were considered at risk for left main obstruction alone, and 3 patients (37.5%) were at risk for double coronary obstruction. All patients underwent successful TAVR without evidence of coronary obstruction. All patients were discharged from the hospital in good clinical condition, and no adverse neurologic events were noted. Procedure success was 100%.
Conclusions:
Evaluation of the first dedicated transcatheter leaflet-splitting device shows that the device can successfully split degenerated bioprosthetic valve leaflets. The procedure was safe and successfully prevented coronary obstruction in patients at risk for this complication following TAVR.