Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Imaging of the human cochlea using micro-computed tomography before and after cochlear implantation: comparison with
Alexandre Karkas1,2,3, Pierre Boureille4,5, Norbert Laroche4
1Service Oto-Rhino-Laryngologie, Division of Otology/Neurotology, Hôpital Nord, University Medical Center of Saint-Etienne, CHU Saint-Etienne, 42055, Saint-Etienne, France. alexandre.karkas@chu-st-etienne.fr.
Purpose:
Analysis of cochlear structures and postoperative temporal bone (TB) imaging are gaining importance in the evaluation of cochlear implantation (CI°). Our aims were to explore the microarchitecture of human cochlea using micro-computed tomography (μCT), analyze electrode's placement inside cochlea after CI°, and compare pre-/post-implantation μCT scans with cone-beam CT (CBCT) scans of same TBs.
Methods:
Cadaveric TBs were scanned using μCT and CBCT then underwent CI° using straight electrodes. Thereafter, they underwent again μCT and CBCT-imaging.
Results:
Ten TBs were studied. μCT allowed visualization of scala tympani, scala vestibuli, basilar membrane, osseous spiral lamina, crista fenestrae, and spiral ligament. CBCT showed same structures except spiral ligament and crista fenestrae. After CI°, μCT and CBCT displayed the scalar location and course of electrode array within the cochlea. There were 7 cases of atraumatic electrode insertion and 3 cases of insertion trauma: basilar membrane elevation, electrode foldover with limited migration into scala vestibuli, and electrode kinking with limited migration into scala vestibuli. Insertion trauma was not correlated with cochlea's size or crista's maximal height but with round window membrane diameter. Resolution of μCT was higher than CBCT but electrode artifacts were similar.
Conclusions:
μCT was accurate in visualizing cochlear structures, and course and scalar position of electrode array inside cochlea with any possible trauma to cochlea or array. CBCT offers a good alternative to μCT in clinical practice for cochlear imaging and evaluation of CI°, with lower radiation and higher resolution than multi-slice CT. Difficulties related to non-traumatic CI° are multifactorial.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...