A Faster Prostate MRI: Comparing a Novel Denoised, Single-Average T2 Sequence to the Conventional Multiaverage T2
Colm B Kelleher1, Jacob Macdonald1, Tracy A Jaffe1
1Duke University Medical Center, Durham, North Carolina, USA.
Journal of Magnetic Resonance Imaging : JMRI
|January 6, 2023
Summary
A denoised, single-average T2-weighted sequence (T2-R) is as effective as the standard multiaverage sequence (T2-S) for detecting prostate cancer lesions and assessing PI-RADS scores. This finding suggests potential for faster prostate MRI examinations without compromising diagnostic accuracy.
Area of Science:
- Radiology
- Medical Imaging
- Oncology
Background:
- Standard T2-weighted (T2-w) sequences in prostate MRI are time-intensive due to the need for multiple signal averages to ensure image quality.
- This time requirement can impact patient comfort and throughput in clinical settings.
Purpose of the Study:
- To evaluate if a denoised, single-average T2-weighted sequence (T2-R) is non-inferior to the conventional multiaverage T2-weighted sequence (T2-S).
- The study aimed to compare lesion detection and Prostate Imaging-Reporting and Data System (PI-RADS) score assessment between the two T2-w sequences.
Main Methods:
- A retrospective analysis of 45 male patients (age 60-75) undergoing prostate MRI was conducted.
- T2-R images were generated from raw data using iterative denoising on a single average, while T2-S used standard multiple averages.
- Nine readers assessed lesion detection and PI-RADS scores, with subsequent side-by-side comparison for subjective preference and image quality evaluation.
Main Results:
- No significant difference was observed in the identification of lesions with PI-RADS ≥3 (P=0.10) or in PI-RADS scoring (P=0.77) between T2-R and T2-S sequences.
- Reader agreement for lesion identification was excellent (ICC=0.84).
- Overall preference for image quality (P=0.07) and sequence preference (ICC=0.62) showed no significant difference, indicating comparable performance.
Conclusions:
- The denoised, single-average T2-weighted sequence (T2-R) is non-inferior to the standard multiaverage T2-weighted sequence (T2-S) for prostate cancer lesion detection and PI-RADS scoring.
- This optimized T2-w sequence offers a potential for reducing MRI scan times without compromising diagnostic efficacy.
- The findings support the clinical utility of accelerated T2-w imaging in prostate MRI protocols.