Preliminary evaluation of a novel secondary check tool for intensity modulated radiotherapy treatment planning
Valeria Casanova Borca1, Lorenzo Radici1, Edoardo Petrucci1
1Medical Physics Department, A.S.L. TO4, 10015 Ivrea (TO), Italy.
Purpose:
To evaluate the dosimetric accuracy of the Delta4 Insight (DI) secondary-check dosimetry system.
Methods:
Absolute dosimetry in reference conditions, output factors, percent depth doses normalized and off-axis dose profiles for different field sizes calculated by DI were compared with measurements. Dose calculations for 20 clinical IMRT/VMAT plans generated in the TPS using both AAA or AcurosXB algorithms were compared with measurements. The average difference between calculated and measured point dose in high-dose region was calculated for all cases. 3D dose measurements were performed in Delta4 Phantom+ and a comparison between calculated and measured dose distributions was performed by means of the gamma analysis with 3 %/2 mm criteria. The dose distributions calculated by DI for 20 IMRT/VMAT plans were compared with those calculated by the TPS.
Results:
The absolute dosimetry computed by DI showed dose value in agreement with the measured one within 0.3 %. The average differences between measured and calculated output factors were less than 2.5 %. The average PDD differences were less than 0.6 %. An excellent agreement between calculations and off-axis measurements is found. The point doses calculated for the 20 recalculated plan showed good agreement with measurements with average differences less than 0.5 %. The average gamma pass rate values for the Delta4 Phantom + 3D dose analysis was greater than 97.%. The comparison of DI with theTPS showed good agreement for the used metrics.
Conclusions:
Delta4 Insight may provide a useful independent secondary dose verification system for IMRT/VMAT plans, complementing the traditional global QA protocols.