Highly selective and sensitive multi-analyte methods for the analysis of steroids are attractive for the diagnosis of endocrine diseases. Commercially available kits are increasingly used for this purpose. These methods involve laborious solid phase extraction, and the respective panels of target analytes are incomplete. We wanted to investigate whether an improvement of kit solutions is possible by introducing automated on-line solid phase extraction (SPE) and combining originally separate analyte panels.
Methods:
Sample preparation was performed using automated on-line SPE on a high-pressure stable extraction column. Chromatographic separation, including isobaric compounds, was achieved using a 0.25 mM ammonium fluoride-methanol gradient on a small particle size biphenyl column. Standard compounds and internal standard mixtures of two panels of a commercially available kit were combined to achieve an optimized and straightforward detection of 15 endogenous steroids. Validation was performed according to the European Medicines Agency (EMA) guidelines with slight modifications.
Results:
Validation was successfully performed for all steroids over a clinically relevant calibration range. Deviations of intra- and inter-assay accuracy and precision results passed the criteria and no relevant matrix effects were detected due to highly effective sample preparation. External quality assessment samples showed the applicability as a routine diagnostic method, which was affirmed by the analyses of anonymized clinical samples.
Conclusions:
It was found possible to complement a commercially available kit for quantitative serum steroid profiling based on isotope dilution LC-MS/MS by implementing automated on-line SPE, thereby improving the practicality and robustness of the measurement procedure.
More Related Videos
10:37Deep Proteome Profiling by Isobaric Labeling, Extensive Liquid Chromatography, Mass Spectrometry, and Software-assisted Quantification
Published on: November 15, 2017
14:39Semi-Targeted Ultra-High-Performance Chromatography Coupled to Mass Spectrometry Analysis of Phenolic Metabolites in Plasma of Elderly Adults
Published on: April 22, 2022
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...