Protein interferes with ionised calcium measurement at the reference electrode liquid junction
Annals of Clinical Biochemistry
|July 1, 1987
Summary
The relationship between ionized calcium and protein is likely due to reference electrode liquid junction potentials, not a Donnan effect. Adjusting electrode solutions revealed this crucial distinction in measurement accuracy.
Area of Science:
- Clinical Chemistry
- Biophysical Chemistry
Background:
- Accurate measurement of ionized calcium (iCa) is critical in clinical diagnostics.
- Previous studies reported a positive correlation between serum protein and iCa using specific analyzers.
- The underlying mechanism for this observed correlation remained unclear, with potential contributions from Donnan effects or electrode artifacts.
Purpose of the Study:
- To investigate the cause of the reported positive relationship between protein concentration and ionized calcium.
- To differentiate between a true Donnan effect and an artifact related to reference electrode liquid junction potentials.
- To assess the impact of reference electrode junction solution tonicity on iCa measurements.
Main Methods:
- Ultrafiltration of human control serum to create sequential retentates with varying protein concentrations.
- Measurement of ionized calcium in retentates and filtrates using Nova 2 and Radiometer ICA1 analyzers.
- Systematic variation of reference electrode liquid junction solutions from hypertonic to isotonic.
Main Results:
- Ionized calcium increased with protein concentration using standard reference electrodes.
- This positive correlation reversed to a slight decrease with increasing protein when isotonic junction solutions were employed.
- Measurements with isotonic solutions paralleled those in the filtrates, suggesting artifactual interference.
Conclusions:
- The observed positive correlation between protein and ionized calcium is primarily an artifact of the reference electrode liquid junction potential.
- Protein's effect on the liquid junction potential, not a Donnan effect, explains the discrepancy in measurements.
- Careful consideration of reference electrode properties is essential for accurate ionized calcium determination in protein-rich samples.