Measurement of the modified TG43 parameters for the bare S7600 Xoft Axxent source model
Autumn E Walter1, Ahtesham U Khan2, Larry A DeWerd2
1Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin, Madison, WI; Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin, Madison, WI.
Purpose:
The purpose of this work is to provide measured data for the modified TG43 parameters [DeWerd et al.] for the newest, Galden-cooled S7600 Xoft Axxent source model.
Methods:
The measurement of radial dose distributions at distances of 1 cm to 4 cm from the source was performed using TLD100 microcubes, EBT3 film, and an Exradin A26 microionization chamber. The overall uncertainty and reproducibility of each dosimeter was evaluated for its use in determining the radial dose function and dose rate conversion coefficient. An acrylic phantom developed in house for previous works was used to measure the polar anisotropy function using TLD100 microcubes at distances of 1 cm, 2 cm, and 5 cm from the source.
Results:
The Exradin A26 chamber was deemed most suitable for measuring the radial dose function. Values determined had a maximum k = 1 uncertainty of 1.4%. The dose rate conversion coefficient measured with the chamber was found to be 9.33 ± 0.21cGy/hrμGy/min. TLD100 microcube measurements of the polar anisotropy had average uncertainties of 6%, 3%, and 2.5% at 1 cm, 2 cm, and 5 cm, respectively.
Conclusions:
The modified TG43 parameters for the bare source were measured with reasonable uncertainty. The values determined will aid with the clinical implementation of the source for breast and endometrial cancer applications.