Effects of core buildup composite resin translucency on intraoral scanner accuracy: an in vitro study
Aim:
The accuracy of 3D images produced by an intraoral scanner (IOS) is affected by the optical characteristics of restorative materials such as metal, ceramic, and composite resin. The present in vitro study aimed to investigate the impact of core buildup composite resin translucency on IOS accuracy.
Materials And Methods:
A core buildup procedure was performed on a proprietary 3D-printed model using injectable composite resins in four groups with different levels of translucency (highest to lowest: AE, A3, AO3, and EX). Ten experimental scans per group were performed using a Medit i700 IOS on a phantom head-mounted model. Reference scans were obtained using an industrial scanner (Solutionix C500). Values of accuracy (trueness and precision) for the respective groups were evaluated using mean deviation values following 3D superimposition.
Results:
Composite resin translucency caused the scale reduction of the optical impressions. Values of trueness showed the highest scale reduction in AE, significantly, followed by A3, AO3, and EX. Considering 50 μm as the cut-off value of deviations for clinical acceptability, the analysis showed most deviations in AE and A3. Similar results were found with precision, where AE showed the highest deviation value statistically, followed by A3, AO3, and EX.
Conclusions:
Composite resin translucency affects the accuracy of optical impressions, causing a fitting error of CAD/CAM prostheses. The more translucent the composite resin, the less accurate the optical impression. This suggests the need for proper compensation during prosthesis designing for an optimal clinical result. In addition, practitioners should indicate in the digital workflow the proper restorative materials regarding not only the mechanical properties and esthetics, but also the optical characteristics.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
X-ray Imaging
Veneer
Other veneering techniques include plain-slicing, quarter-slicing, and rift-slicing. These...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT