Analytical Performance Evaluation for Estradiol using Liquid Chromatography-Tandem Mass Spectrometry
Eun Jeong Won1, Ahram Yi2, Young Jin Ko3
1Department of Laboratory Medicine, Chonnam National University Hwasun Hospital, Hwasun, South Korea.
Background:
There is an increased need for the sensitive and accurate measurement of estradiol levels in patients with estradiol-related endocrine disorders. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is a reliable and accurate method for measuring steroid hormone levels. Here, we aimed to establish an LC-MS/MS-based method to quantify estradiol levels without sample derivatization and evaluated its analytical performance.
Methods:
Sciex Triple Quad 6500 + LC-MS/MS was used for estradiol analysis. We evaluated its analytical performance, including linearity, precision, the lower limit of detection and quantification (LoD and LoQ, respectively), accuracy, and carryover. The estradiol results determined by LC-MS/MS were compared with those obtained using a chemiluminescent microparticle immunoassay.
Results:
The LC-MS/MS output was linear for serum estradiol concentrations in the range of 0.2-10311.6 pmol/L. The intra-laboratory precision (coefficient of variation) was 3.0-10.1 %. The LoD and LoQ were 2.8 and 7.5 pmol/L, respectively. The overall accuracy was within 15 % of bias, and the carryover was within the acceptable range (<1.0 %). The results of the estradiol analysis determined by LC-MS/MS were comparable to those obtained by the chemiluminescent microparticle immunoassay (r2 = 0.9843), although there was a negative bias of - 17.82 (95 % confidence interval, -27.21 to - 8.44).
Conclusions:
A highly sensitive, derivatization-free LC-MS/MS method was successfully developed in this study. This may be beneficial for estradiol measurements in patients with estradiol-related endocrine disorders.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
07:26A Simple Method for Automated Solid Phase Extraction of Water Samples for Immunological Analysis of Small Pollutants
Published on: January 1, 2016
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process