Feasibility of diffuser-incorporated near-infrared trans-reflectance measurement for quantitative detection of
Yunjung Kim1, Sanghoon Cho1, Hoeil Chung1
1Department of Chemistry and Research Institute for Convergence of Basic Science, Hanyang University, Seoul, 04763, Republic of Korea.
Abstract:
A diffuser-incorporated near-infrared (NIR) trans-reflectance measurement is demonstrated for quantitative detection of polyethylene (PE) particles captured in perfluorohexane (PFH, C6F14). PFH effectively captures PE particles through its hydrophobicity and absorbs little NIR radiation, recommending it for use in background-free NIR detection of captured PE particles. A reflective metal disk was used to push the captured PE particles in PFH toward the bottom of the vial that contained the sample, and the trans-reflectance measurement was performed by illuminating NIR radiation from the bottom of the vial at 45o. Reproducibility is limited by the variation in the positions of small PE-particle aggregates at the water/PFH interface and the difficulty in ensuring full NIR sampling (coverage) of large aggregates. An effective way to secure improved reproducibility under these circumstances is illumination of broader and more uniform NIR radiation for measurement. For this purpose, a polytetrafluoroethylene (PTFE) disk was uniquely incorporated as a diffuser for the trans-reflectance measurement. Compared to the measurement with no diffuser, the diffuser-incorporated scheme produced more distinct PE peaks of low-quantity samples (0.1 and 0.2 mg) and enhanced the reproducibility in measurements of all the samples (0.1-4.0 mg of PE). As a result, the correlation between peak intensity and particle quantity was excellent (R2: 0.997), and a limit of detection of 0.07 mg was achieved.